Do this...
16.6
Decrease the output to 40mA.
16.7
Increase the output to 60mA.
Decrease the rate to 30 ppm.
16.8
Increase the rate to 180ppm.
16.9
Connect the ECG cable to the E Series and
QED 6 analyzer. Select the ECG at 60 BPM on
the QED 6 analyzer. Decrease the pacer rate
on the unit to 58ppm.
16.10
Press the Async Pace softkey.
16.11
Turn off the QED 6 analyzer. Set Pacer Rate to
100ppm. Press the RECORDER ON button.
16.12
Press and hold 4:1 button.
Record your results on the Maintenance Tests Checklist.
1-28
Verify that...
Output on the QED 6 analyzer is 40mA or +/- 5mA. Pulse width is 
40mS +/-2mS.
Pacer rate on QED 6 analyzer is 29-31 ppm
Pacer rate on QED 6 analyzer is 177-183 ppm
The unit displays ECG at 60 BPM with no stimulus markers.
The unit displays ECG at 60 BPM with stimulus markers, and displays the
Async pace message.
The unit displays pace stimulus markers every 15mm +/-1mm.
The unit displays pace stimulus markers every 60 mm+/- 1.5 mm.
Pass/Fail
o o
o o
o o
o o
o o
o o
o o