Error Message
CONFIRM MANUAL MODE
DEFIB DISABLED
DEFIB FAULT 71
DEFIB FAULT 72
DEFIB FAULT 76
DEFIB FAULT 77
DEFIB FAULT 78
DEFIB FAULT 79
DEFIB FAULT 80
DEFIB FAULT 81
DEFIB FAULT 84
DEFIB FAULT 85
DEFIB FAULT 86
DEFIB FAULT 87
DEFIB FAULT 94
DEFIB FAULT 95
DEFIB FAULT 108
9650-1450-01 Rev. D
Explanation
Displayed when manual mode is entered. Alerts user to
confirm that manual mode is desired.
User prompt issued simultaneously with other faults if
defib is disabled.
More than 50 internal dumps occurred in less than 20
minutes.
Charging inhibited, voltage out of range.
Capacitor voltage too high for selected energy.
Capacitor voltage > than absolute rated max.
Unable to charge defib cap.
Defibrillator charging too slowly.
4 defibrillator faults detected within 20 second period.
Discharge switch in undefined state.
"Upper" discharge transistor shorted (measured via
applicable A/D channel).
"Lower" discharge transistor shorted (measured via
applicable A/D channel).
One discharge switch closed during power up test.
Both discharge switches closed during power up test.
Processor fault causing safety monitor port to be non-
functional.
Safe or shutdown line is not functional.
VMON voltage is less than the target energy during
charging.
User
Technical Action
Advisory
Possible configuration problem. Replace high
voltage module. Call ZOLL Technical Support.
Turn the unit to OFF and back on. If fault persists,
replace high voltage module.
Turn the unit to OFF and back on. If fault persists,
replace high voltage module.
Replace high voltage module or capacitor.
Replace high voltage module or capacitor.
Replace high voltage module or capacitor.
Replace high voltage module or capacitor.
Replace high voltage module or capacitor.
Replace high voltage module or capacitor.
Replace high voltage module.
Replace high voltage module.
Replace paddles, control board or system board.
Replace paddles, control board or system board.
Replace system board, high voltage module or
capacitor.
Replace high voltage module.
Replace high voltage module or capacitor.