RIGOL

To Test Continuity

In view of Continuity measurement function, the following part demonstrated how to
link the measurement connection and how to choose measurement functions. The
following practice will gradually guide you to be familiar with the Continuity
measurement technique.
Table 1-8 Continuity measurement characteristics
Tests Current
Max Resolution
Open-circuit Voltage
Import Protection
Configurable Parameters
Basic measurement:
1. Connect test leads as Figure 1-28 shown. Red test lead connects the HI Terminal,
Black test lead connects the LO Terminal.
2. Press
to select the Continuity Measurement
3. Setup the Short-circuit resistance.
Press
Res , to set up the Short-circuit Impedance. Default value of the
Short-circuit Impedance will be 10MΩ, this parameter had been setup, and the user
may carry on the Continuity measurement directly without modification.
© Copyright RIGOL Technologies, Inc. 2007.
Figure 1-27
1mA
Range fixed at 2KΩ
<5V
1000V (HI Terminal)
0≤R
≤Short-circuit impedance
testing
(0Ω≤Short-circuit impedance≤2kΩ)