4 Installation
4.2 Hardware installation

4.2.2 Connect the system components

4.2.2
Connect the system components
Connect system units to the
rocker
ReadyToProcess Pump 25 and ReadyToProcess CBCU units are connected to the rocker
with UniNet cables. Power to the units is supplied from the rocker.
Follow the instructions below to connect the system units to the rocker.
See Rear view of the rocker, on page 46 for the location of connectors on the rear panel
of the rocker.
Step
1
86
NOTICE
If UniNet cables are disconnected during a run, the run will be in-
terrupted and must be restarted. Make sure that cables are prop-
erly secured before starting a run and avoid moving system units
during a run.
Action
Connect the ReadyToProcess CBCU unit(s) to the rocker.
1 Connect the supplied UniNet cable between the UniNet-9 port on the
CBCU rear panel (see Rear view of ReadyToProcess CBCU, on page 53)
and the first available UniNet-9 port on the rocker rear panel. Up to two
ReadyToProcess CBCU units are used for operation in dual mode.
2 For single mode, set the CAN ID switch on the CBCU to position 1. For
dual mode using two CBCU units, set the CAN ID switches on the two
CBCU units to positions 1 and 2 respectively.
Tip:
In dual mode, label the CBCU units with their respective CAN ID to simplify
identification in the UNICORN software.