7 Troubleshooting

7.1 ReadyToProcess WAVE 25 system

7.1

ReadyToProcess WAVE 25 system

Alarm messages
For many of the problems that may occur, UNICORN displays an alarm message on the
screen. Follow the instruction shown to resolve the problem.
System units not recognized
Error symptom
CBCU not recognized
by UNICORN.
Pump not recognized
by UNICORN.
CBCU and/or pump not
recognized by
UNICORN.
156
Possible cause
Incorrect CBCU CAN ID.
CBCU is not selected as
a component.
Incorrect pump CAN ID.
Pump is not selected as
a component.
Gap between occupied
UniNet-9 ports on the
rocker rear panel.
One or more connected
pumps is not defined in
UNICORN.
Corrective action
Check and if necessary change CAN ID using the
switch on the CBCU rear panel. See CAN ID, on
page 54 for details. The system must be switched
off before the CAN ID is changed, and then
restarted.
Click Edit in System Properties in the Administra-
tion module and add the CBCU as a component.
Check and if necessary change CAN ID using the
switch on the pump rear panel. See CAN ID, on
page 57 for details. Each pump must have a
unique CAN ID value. The system must be switched
off before the CAN ID is changed, and then
restarted.
Click Edit in System Properties in the Administra-
tion module and add the pump as a component.
Reposition the connectors or insert jumpers in the
unused positions between the occupied ports.
See Connect system units to the rocker, on page 86.
Read the displayed message when you connect
to the system in System Control. Make sure all
available components are selected for the system.
See Section 5.2.3 Configure system properties, on
page 127.
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD