Analog Output Trim; Table 7- 10 Analog Out Trim Menu Structure - Siemens SITRANS F Series Operating Instructions Manual

Ultrasonic flowmeter
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Log Memory Left
This view only menu cell shows the minimum remaining number of characters available for
Datalogger and site storage.
Defragment
Selecting YES for this item consolidates memory data blocks into contiguous storage;
collapsing the filler regions. You may be able to use an additional block for site or Datalogger
storage as a result. Use this command if you seem to be out of memory even though the
[Log Memory Left] item indicates free capacity.
To access the [Memory Control] menu cells:
1. From the [Meter Type] menu scroll down to the [Meter Facilities ] menu
2. Press the <Right Arrow> key and then scroll to the [Memory Control] menu.
3. Press the <Right Arrow> key to access the [Log Memory Left] menu cell and view the
4. Press the <Down Arrow> key to select the [Defragment] menu.
5. To select [Yes] press the <Right Arrow> and then <ENTER> key.
7.10

Analog Output Trim

Introduction
Analog Out Trim function allows fine-tuning of the flow meter's analog current output using
an ammeter connected to the output under test.
Table 7- 10
Analog Out Trim
FST020 IP65 NEMA 4X
Operating Instructions, 02/2012, A5E03086487-05
data memory bytes available.
Analog Out Trim Menu Structure
Io1
7.10 Analog Output Trim
Operate
Trim@4mA Indicated mA=x.xx
Functions
91