Programme settings
Programme
header
98
Adjusting programme settings
The programme settings should be adjusted to suit technical
requirements and the load.
Additional specialist knowledge is required to alter programme
settings and this should therefore be undertaken only by experienced
users or by Miele Service.
Changing programme parameters on a validated machine will
necessitate a renewed performance validation.
Programme structure
Each programme is subdivided into programme blocks which run one
after another. A programme consists of at least one and a maximum
of eleven programme blocks. Each block can occur only once in a
programme.
The so-called programme header is placed above the programme
blocks and contains general programme settings. Individual wash
block parameters are also globally activated or deactivated here.
Spray arm monitoring
–
It is possible to monitor spray arm rotation in selected wash blocks.
Change volume of water
–
The water intake quantity can be increased or reduced in each
programme. The setting is then valid for all programme blocks
including water intake.
Drain time
–
If the on-site drainage system is insufficient to drain the waste
water from the wash cabinet within the time allocated, the drainage
time can be increased.