Technical data
Height with machine lid
Height without machine lid
Width
Depth
Depth with door open
Wash cabinet dimensions:
height
width
depth of upper basket/lower basket
Weight (net)
Max. load capacity of open door
Voltage, connected load, fuse rating
Mains cable
Water temperature water connection:
Cold water / Steam condenser
Hot water / AD water (optional)
Static water pressure
Minimum water connection flow pressure:
Cold water / steam condenser
Hot water
AD water (optional)
Recommended water connection flow pressure:
Cold water / hot water
AD water (optional)
Steam condenser
AD water connection without pressure (optional)
Delivery head
Drainage length
Operation:
Ambient temperature
Relative humidity maximum
linear decreasing to
Storage and transportation conditions:
Ambient temperature
Relative humidity
Air pressure
Altitude above sea level
Protection category (according to IEC 60529)
Soiling level (according to IEC/EN 61010-1)
Overvoltage category (according to IEC 60664)
Noise level in dB (A),
sound pressure LpA during cleaning and drying phases
Test certificates
mark
Manufacturer's address
* If installed above 1,500 m the boiling point of water will be lower. In this case the disinfecting temperature and the
holding time will need to be reset.
184
835 mm
820 mm
598 mm
598 mm
1,200 mm
520 mm
530 mm
474 mm/520 mm
74 kg
37 kg
See data plate
Approx. 1.8 m
max. 20 °C
max. 65 °C (UK: max. 60 °C)
Max. 1,000 kPa pressure
100 kPa (UK: 100 kPa) pressure
40 kPa (UK: 100 kPa) pressure
30 kPa (UK: 100 kPa) pressure
≥ 200 kPa pressure
≥ 200 kPa pressure
≥ 100 kPa pressure
8.5-60 kPa
min. 0.3 m, max. 1.0 m
max. 4.0 m
5 °C to 40 °C
80 % for temperatures up to 31 °C
50 % for temperatures up to 40 °C
- 20 °C to 60 °C
10 % to 85 %
500 hPa to 1060 hPa
to 1,500 m*
IP21
2
II
< 70
VDE, EMC radio suppressed
2006/42/EC Machinery directive
Miele & Cie. KG, Carl-Miele-Straße 29,
33332 Gütersloh, Germany