Miele PG 8583 Operating Instructions Manual page 125

Washer-disinfector for laboratory glassware and laboratory utensils
Table of Contents

Advertisement

Test programmes
Various programmes are available for monitoring cleaning
performance in routine testing. The test programmes are not separate
processing programmes. Rather, they are additional functions that
can be activated prior to starting any processing programme.
The test programmes interrupt the programme sequence
automatically at specified points. The interruption is indicated by an
audible signal tone and message on the display. Miele Service can
set the duration of the interruption to between 10 seconds and
approx. 42 min. During this time period, measurements can be made
or the door can be opened to obtain a sample.
To prevent cooling of the wash cabinet, do not keep the door open
too long.
After the time period elapses, the programme sequence continues
automatically. If the door has been opened, the programme cannot
resume until the door has been closed again.
If a measurement or sample is not needed, you can resume the
programme sooner by pressing the Start/Stop button.
The following test programmes can be selected:
Laboratory
The programme sequence can be paused in each wash block
immediately before the wash fluid is drained away.
Validation
The programme sequence is interrupted at the following points:
– before the wash fluid is drained away in the final wash block,
– after the interim rinse before the wash fluid is drained away, and
– after water intake and before draining in the final rinse block.
Maintenance
125