Cover plate/
Worktop
protector
A cover plate/worktop protector is available from Miele and may be
advisable, depending on building-in location.
The cover plate protects the lower edge of the worktop from steam
damage.
Electromagnetic compatibility (EMC)
The washer-disinfector has been tested for electromagnetic
compatibility in accordance with EN 61326-1 and is suitable for
operation in commercial environments, such as hospitals, medical
practices and laboratories and other similar environments which are
connected to the mains power supply.
The washer-disinfector's HF emissions are very low and are therefore
unlikely to interfere with other electronic appliances in the vicinity.
Flooring in the installation area must be wood, concrete or tiled.
Synthetic flooring must be able withstand a relative humidity level of
30 % to minimise the risk of electrostatic discharges.
The quality of the power supply should comply with that found in a
typical commercial or hospital environment and should deviate from
the nominal voltage by a maximum of +/- 10 %.
Installation
149