18
Averaging
Purpose
Use this program to determine weights under extreme-
ly unstable ambient conditions. In this program, the
balance/scale calculates the weight as the average
value from a defined number of individual weighing
operations. These weighing operations are also
known as "subweighing operations" or "subweighs."
Available Features
– The measured result displayed is the arithmetic
mean shown in the selected weight unit;
a triangle under the plus or minus sign indicates
that this is a calculated value
– You can set the number of subweighing
operations performed in the operating menu:
See "Configuring the Balance/Scale"
– Press v for at least 2 sec. to display the pre-set
number of subweighing operations
– Press v to toggle the display between the
calculated result and the weight
Factory Settings
Number of subweighs for averaging: 10 (3. 3. 2)
Preparation
§ Configure the Averaging application:
See "Configuring the Balance/Scale"
Menu code 2. 1.12
– Number of subweighs for averaging:
5 subweighs
3. 3. 1
10 subweighs
3. 3. 2
20 subweighs
3. 3. 3
50 subweighs
3. 3. 4
100 subweighs
3. 3. 5
See also "Configuring the Balance/Scale"