Averaging - Sartorius Basiclite Installation And Operating Instructions Manual

Electronic analytical and precision balances and gold scales with ec type approval
Table of Contents

Advertisement

18

Averaging

Purpose
Use this program to determine weights under extreme-
ly unstable ambient conditions. In this program, the
balance/scale calculates the weight as the average
value from a defined number of individual weighing
operations. These weighing operations are also
known as "subweighing operations" or "subweighs."
Available Features
– The measured result displayed is the arithmetic
mean shown in the selected weight unit;
a triangle under the plus or minus sign indicates
that this is a calculated value
– You can set the number of subweighing
operations performed in the operating menu:
See "Configuring the Balance/Scale"
– Press v for at least 2 sec. to display the pre-set
number of subweighing operations
– Press v to toggle the display between the
calculated result and the weight
Factory Settings
Number of subweighs for averaging: 10 (3. 3. 2)
Preparation
§ Configure the Averaging application:
See "Configuring the Balance/Scale"
Menu code 2. 1.12
– Number of subweighs for averaging:
5 subweighs
3. 3. 1
10 subweighs
3. 3. 2
20 subweighs
3. 3. 3
50 subweighs
3. 3. 4
100 subweighs
3. 3. 5
See also "Configuring the Balance/Scale"