g. Fault 7, High Case Temp
The HEATER LED flashes three times,
pauses for three seconds, and repeats (Fig-
ure 5-7).
1. If the case temperature exceeds 185°F
(85°C), the temperature control will
shut off and a fieldbus alarm will be
sent.
2. This signifies that the environment
where the Oxymitter 5000 is installed
exceeds the ambient temperature re-
quirements or that heat due to convec-
tion is causing case temperature to rise
above the limit.
3. Placing a spool piece between the
stack flange and the Oxymitter 5000
flange may eliminate this problem.
4. If a spool piece does not solve the
problem, relocation is the only solution.
IB-106-350 Rev. 1.2
April 2001
Troubleshooting
5-9