Handling the Oxymitter 5000
It is important that printed circuit
boards and integrated circuits are
handled only when adequate antistatic
precautions have been taken to pre-
vent possible equipment damage.
The Oxymitter 5000 is designed for in-
dustrial applications. Treat each com-
ponent of the system with care to
avoid physical damage. Some probe
components are made from ceramics,
which are susceptible to shock when
mishandled.
g. System Considerations
Prior to installing your Oxymitter 5000,
make sure you have all the components
necessary to make the system installation.
Ensure all the components are properly in-
tegrated to make the system functional.
After verifying that you have all the compo-
nents, select mounting locations and deter-
mine how each component will be placed in
terms of available line voltage, ambient
temperatures, environmental considera-
tions, convenience, and serviceability.
Figure 1-3 shows a typical system wiring. A
typical system installation is illustrated in
Figure 1-4.
2 CALIBRATION GAS LINES
[300 FT (90 M) MAX]

Figure 1-3. Oxymitter 5000 FOUNDATION Fieldbus Connections

1-4
Description and Specifications
OXYMITTER 5000
WITH INTEGRAL ELECTRONICS
BY CUSTOMER
LINE VOLTAGE
A source of instrument air is optional at the
Oxymitter 5000 for reference air use. Since
the unit is equipped with an in-place cali-
bration feature, provisions can be made to
permanently connect calibration gas tanks
to the Oxymitter 5000.
If the calibration gas bottles will be perma-
nently connected, a check valve is required
next to the calibration fittings on the integral
electronics. This check valve is to prevent
breathing of the calibration gas line and
subsequent flue gas condensation and cor-
rosion. The check valve is in addition to the
stop valve in the calibration gas kit or the
solenoid valves in the IMPS 4000 or SPS
4000.
The integral electronics is rated NEMA
4X (IP66) and is capable of operation
at temperatures up to 149°F (65°C).
Retain the packaging in which the
Oxymitter 5000 arrived from the fac-
tory in case any components are to be
shipped to another site. This packag-
ing has been designed to protect the
product.
1-3

IMPS 4000 (OPTIONAL)

Information on the IMPS 4000 is available in the
IMPS 4000 Intelligent Multiprobe Test Gas Se-
quencer Instruction Bulletin.
FIELDBUS DIGITAL
SIGNAL
FIELDBUS COMPUTER
Rosemount Analytical Inc.
A Division of Emerson Process Management
Oxymitter 5000
NOTE
TERMINAL
28550005