Emerson Oxymitter 5000 Instruction Manual page 130

Oxygen transmitter with foundation fieldbus communications
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Model 4081FG Oxygen Analyzer with F
Parameter
Index
Number
BLOCK_ALM
21
BLOCK_ERR
06
CHANNEL
15
FIELD_VAL
19
GRANT_DENY
12
HI_ALM
34
HI_HI_ALM
33
HI_HI_LIM
26
HI_HI_PRI
25
HI_LIM
28
HI_PRI
27
IO_OPTS
13
L_TYPE
16
LO_ALM
35
LO_LIM
30
LO_LO_ALM
36
LO_LO_LIM
32
LO_LO_PRI
31
LO_PRI
29
LOW_CUT
17
MODE_BLK
05
OUT
08
OUT_D
37
OUT_SCALE
11
PV
07
B-2
OUNDATION
Units
None
The block alarm is used for all configuration, hardware, connection failure or system
problems in the block. The cause of the alert is entered in the subcode field. The first alert
to become active will set the Active status in the Status parameter. As soon as the
Unreported status is cleared by the alert reporting task, another block alert may be reported
without clearing the Active status, if the subcode has changed.
None
This parameter reflects the error status associated with the hardware or software
components associated with a block. It is a bit string, so that multiple errors may be shown.
None
The CHANNEL value is used to select the measurement value. Refer to the appropriate
device manual for information about the specific channels available in each device.
You must configure the CHANNEL parameter before you can configure the XD_SCALE
parameter.
Percent
The value and status from the transducer block or from the simulated input when simulation
is enabled.
None
Options for controlling access of host computers and local control panels to operating,
tuning, and alarm parameters of the block. Not used by device.
None
The HI alarm data, which includes a value of the alarm, a timestamp of occurrence and the
state of the alarm.
None
The HI HI alarm data, which includes a value of the alarm, a timestamp of occurrence and
the state of the alarm.
EU of PV_SCALE
The setting for the alarm limit used to detect the HI HI alarm condition.
None
The priority of the HI HI alarm.
EU of PV_SCALE
The setting for the alarm limit used to detect the HI alarm condition.
None
The priority of the HI alarm.
None
Allows the selection of input/output options used to alter the PV. Low cutoff enabled is the
only selectable option.
None
Linearization type. Determines whether the field value is used directly (Direct), is converted
linearly (Indirect), or is converted with the square root (Indirect Square Root).
None
The LO alarm data, which includes a value of the alarm, a timestamp of occurrence and
the state of the alarm.
EU of PV_SCALE
The setting for the alarm limit used to detect the LO alarm condition.
None
The LO LO alarm data, which includes a value of the alarm, a timestamp of occurrence and
the state of the alarm.
EU of PV_SCALE
The setting for the alarm limit used to detect the LO LO alarm condition.
None
The priority of the LO LO alarm.
None
The priority of the LO alarm.
%
If percentage value of transducer input fails below this, PV = 0.
None
The actual, target, permitted, and normal modes of the block.
Target: The mode to "go to"
Actual: The mode the "block is currently in"
Permitted: Allowed modes that target may take on
Normal: Most common mode for target
EU of OUT_SCALE
The block output value and status.
None
Discrete output to indicate a selected alarm condition.
None
The high and low scale values, engineering units code, and number of digits to the right of
the decimal point associated with OUT.
EU of XD_SCALE
The process variable used in block execution.
fieldbus Communications
Description