Figure 4-8. Heater Strut Assembly - Emerson Oxymitter 5000 Instruction Manual

Oxygen transmitter with foundation fieldbus communications
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Oxymitter 5000
V-DEFLECTOR
CERAMIC
DIFFUSER
CELL FLANGE
ASSEMBLY
d. Loosen, but do not remove, the three
screws (34, Figure 4-1) on the strut in the
finned housing. The spring tension should
release and the strut moves up.
e. Grasp the wire loop and carefully slide the
strut out of the probe tube (Figure 4-8).
f.
When replacing the strut, align the slot on
the heater plate with the calibration gas line
in the probe tube. Slide the strut into the
probe tube. It will turn to align the hole on
the back plate of the strut with the calibra-
tion gas line. When the hole and the cali-
bration gas line are aligned correctly, the
strut will slide in the rest of the way.
Rosemount Analytical Inc.
A Division of Emerson Process Management
CERAMIC SUPPORT ROD
HEATER

Figure 4-8. Heater Strut Assembly

Instruction Manual
g. Push down on the back plate of the strut to
make sure you have spring tension and
then tighten the three screws on the back
plate.
h. Replace the CAL and REF gas silicon
tubes.
i.
Install the entire electronics per paragraph
4-5a, steps 7 through 13.
j.
Follow the instructions in paragraph 4-4a.2
to install the Oxymitter 5000 into the stack
or duct. If installing an Oxymitter 5000/SPS
4000 assembly, follow the instructions in
paragraph 4-4b.2.
Maintenance and Service
IB-106-350 Rev. 1.2
April 2001
WIRE
LOOP
22220050
4-13