LED STATUS INDICATORS

a. Diagnostic/Unit Alarms
Table 4-1 lists the types and status of
alarms that will be encountered. (See
Section 5, TROUBLESHOOTING, for a de-
tailed description of each fault).
b. When the electronics determines a calibra-
tion is recommended, the CALIBRATION
RECOMMENDED LED is on solid.
c. The CAL LED turns on when a calibration is
recommended and is on during the calibra-
tion process. During calibration, the CAL
LED can be flashing, which would indicate
operator action is requested, or on solid,
which indicates calculations and measure-
ments are in progress.

Table 4-1. Diagnostic/Unit Alarms

LED
FLASHES
1
HEATER T/C
2
3
4
1
2
HEATER
3
4
5
1
O
CELL
3
2
4
CALIBRATION
1
2
3
4-6
Maintenance and Service
STATUS
FAULT
OPEN
1
SHORTED
2
REVERSED
3
A/D COMM
4
ERROR
OPEN
5
HIGH HIGH
6
TEMP
HIGH CASE
7
TEMP
LOW TEMP
8
HIGH TEMP
9
HIGH mV
10
BAD
11
EEPROM
12
CORRUPT
INVALID
13
SLOPE
INVALID
14
CONSTANT
LAST
15
CALIBRATION
FAILED
It is recommended that the Oxymitter
5000 be removed from the stack for all
service activities. The unit should be
allowed to cool and be taken to a clean
work area. Failure to comply may
cause severe burns.
Disconnect and lock out power before
working on any electrical components.
There is voltage up to 115 VAC.
4-4
OXYMITTER 5000 REMOVAL/
REPLACEMENT
a. Oxymitter 5000 (without Integrally
Mounted SPS 4000)
1. Remove.
(a) Turn off power to the system.
(b) Shut off the calibration gases at the
cylinders and the instrument air.
(c) Disconnect the calibration gas and
instrument air lines from the Oxy-
mitter 5000.
(d) While facing the Oxymitter 5000
and looking at the Rosemount la-
bel, remove screw (36, Figure 4-1),
gasket (37) and cover lock (38) se-
curing left housing cover (31). Re-
move the cover to expose the
terminal block Figure 4-4.
(e) Loosen the screw on the AC termi-
nal cover and slide the cover back
to access the neutral and line ter-
minals. Loosen the AC line and
neutral terminal screws and re-
move the leads. Loosen the ground
lug screws and remove the leads.
Slide the line power leads out of
the AC line voltage port.