Figure 5-12. Fault 12, Eeprom Corrupt - Emerson Oxymitter 5000 Instruction Manual

Oxygen transmitter with foundation fieldbus communications
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Instruction Manual
IB-106-350 Rev. 1.2
April 2001
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC
INC
HIGH
LOW
GAS
GAS
DEC
DEC

Figure 5-12. Fault 12, EEPROM Corrupt

5-14
Troubleshooting
HEATER T/C
HEATER
SW2
O2 CELL
CALIBRATION
J1
O2 CELL mV +
TP1
O2 CELL mv -
TP2
HEATER T/C +
TP3
HEATER T/C -
TP4
CAL
TEST GAS +
TP5
PROCESS -
TP6
% O2
28550024
l.
Fault 12, EEPROM Corrupt
The O
CELL LED flashes four times,
2
pauses for three seconds, and repeats (Fig-
ure 5-12).
1. This alarm can occur if the EEPROM is
changed for a later version. At power
up, the EEPROM is not updated.
2. To correct this problem, power down
and then restore power. The alarm
should clear.
3. If the alarm occurs while the unit is
running, there is a hardware problem
on the microprocessor board.
4. If cycling the power does not clear the
alarm, see paragraph 4-5b, Electronic
Assembly Replacement.
Rosemount Analytical Inc.
A Division of Emerson Process Management
Oxymitter 5000