Cell Replacement; Figure 4-9. Cell Replacement Kit - Emerson Oxymitter 5000 Instruction Manual

Oxygen transmitter with foundation fieldbus communications
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Instruction Manual
IB-106-350 Rev. 1.2
April 2001
PROBE TUBE
(NOT INCLUDED
IN KIT)
CALIBRATION GAS
PASSAGE

Figure 4-9. Cell Replacement Kit

4-8

CELL REPLACEMENT

This paragraph covers oxygen sensing cell re-
placement. Do not attempt to replace the cell
until all other possibilities for poor performance
have been considered. If cell replacement is
needed, order the cell replacement kit
(Table 8-1).
The cell replacement kit (Figure 4-9) contains a
cell and flange assembly, corrugated seal,
setscrews, socket head cap screws, and anti-
seize compound. The items are carefully pack-
aged to preserve precise surface finishes. Do
not remove items from the packaging until they
are ready to be used. Spanner wrenches and
hex wrenches needed for this procedure are
part of an available special tools kit (Table 8-1).
4-14
Maintenance and Service
CORRUGATED
SEAL
CELL AND
FLANGE
ASSEMBLY
SOCKET HEAD
CAP SCREWS
22220028
Use heat-resistant gloves and clothing
when removing the probe. Do not at-
tempt to work on these components
until they have cooled to room tem-
perature. Probe components can be as
hot as 800°F (427°C). This can cause
severe burns.
Disconnect and lock out power before
working on any electrical components.
There is voltage of up to 115 VAC.
Do not remove the cell unless certain
it needs to be replaced. Removal may
damage the cell and platinum pad. Go
through the complete troubleshooting
procedure to make sure the cell needs
to be replaced before removing it.
a. Follow the instructions in paragraph 4-4a.1
to remove the Oxymitter 5000 from the
stack or duct. If removing an Oxymitter
5000/SPS 4000 assembly, follow the in-
structions in paragraph 4-4b.1.
b. If the probe uses the standard diffusion
element, use a spanner wrench to remove
the diffusion element.
Rosemount Analytical Inc.
A Division of Emerson Process Management
Oxymitter 5000