Sample Programming and Processing
6.2 Prepare Racks for Analysis
6.2 Prepare Racks for Analysis
This section describes how to load racks on the rack set unit and priority rack set unit when the
AU5800 is connected to the Beckman laboratory automation system. It is necessary to load reagent
blank (blue), calibration (yellow), and QC (green) racks on the rack set unit or priority rack set unit.
Routine (white) and emergency (red) racks can be loaded on the rack set unit and priority rack set
unit as needed.
6.2.3 Place a Set Tray and Collection Tray on the Analyzer
One tray can be placed on the rack set unit, and one tray can be placed on the rack collection unit.
To place trays:
1
Confirm that the rack set unit amber LED and rack collection unit amber LED are not blinking.
Figure 6.1
1. Rack set unit amber LED
2. Rack collection unit amber LED
3. Rack stabilizing bar
WARNING
Before removing a tray from the system during operation, confirm that the amber
LED is off. If a tray is removed while the amber LED is blinking, an injury may result
or the rack may topple over.
For a detailed description of the amber LEDs, refer to
Feeder Unit)
6-2
1
in chapter 3.
3
2
3.4.3 LEDs and Operation Buttons (Rack
PN B08501AB