Cautions - Nonin RespSense LS1R-9R Operator's Manual

Capnography monitor
Table of Contents

Advertisement

Cautions

RespSense should only be operated by trained licensed practitioners.
To prevent damage to the monitor, operate and store the monitor in an upright position.
Setting alarm limits to extremes can render the alarm system useless.
To ensure the longevity and safety of the monitor, the carrying case must be used when operating the monitor in
an emergency medical services (EMS) environment.
Secure RespSense with mounting hardware if used in transport vehicles.
Do not mount RespSense directly above the patient. If the monitor is mounted, be sure to check that the
adjustable mounting clamp is securely affixed.
When mounting the monitor to a mobile pole, mounting the monitor higher than 1.5 meters (5 feet) or mounting
more than 2 kilograms (4.5 pounds) of equipment onto the pole may result in tipping, damage to the equipment,
or injury.
Always turn off the monitor prior to cleaning the monitor or changing the moisture trap and/or filter.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Do not sterilize or autoclave the monitor or accessories. Do not immerse in liquids.
This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment
and/or systems. This standard is designed to provide reasonable protection against harmful interference in a
typical medical installation. However, because of the proliferation of radio-frequency transmitting equipment and
other sources of electrical noise in healthcare and other environments, it is possible that high levels of such
interference due to close proximity or strength of a source might disrupt the performance of this device. Medical
electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into
service according to the EMC information specified in this manual.
Each time the system is turned on, alarms are silenced for 2 minutes unless the operator presses the Audible
Alarm Pause/Resume button.
Set or adjust alarm parameters one at a time.
Do not cover or block speaker opening. This may significantly reduce the sound volume.
Before each use, it is the operator's responsibility to verify the alarm limits are appropriate for the patient being
monitored.
The patient's nasal passage may dry out if continuous monitoring is required. Check patient hourly for nasal
comfort.
If the EtCO
value is out of normal range (4.4 – 5.7 Vol%/kPa or 33 – 43 mmHg) an internal air leak is possible.
2
Replace the moisture trap and repeat the calibration procedure. If the problem persists, contact Nonin Technical
Service.
If RespSense is intended to be stored for longer periods of time, always charge the battery to full capacity before
storing it in order to prevent damage to the equipment.
Avoid rapid temperature change or extreme temperatures. This can cause malfunction.
Never store or transport RespSense where condensation can occur. However, if this has occurred, wait until all
condensation has evaporated before using RespSense.
2

Advertisement

Table of Contents
loading

Table of Contents