Cautions - Nonin WristOx2 3150 Operator's Manual

Hide thumbs Also See for WristOx2 3150:
Table of Contents

Advertisement

Warnings (Continued)
Only use Nonin-branded sensors with a length of 1 meter or less. Accuracy may degrade if sensor cable is over
1 meter in length. Using the sensor cable adapter does not affect accuracy.
Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the ME system,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.

Cautions

If this device fails to respond as described, refer to "Troubleshooting" or discontinue use until the situation has
been corrected. Contact Nonin Technical Service.
This device has motion tolerant software that minimizes the likelihood of motion artifact being misinterpreted as
good pulse quality. In some circumstances, however, the device may still interpret motion as good pulse quality.
Do not place liquids on top of this device.
When setting the clock in Programmed Activation Mode using nVISION software, verify all set times and dates
are valid.
Do not place the WristOx
2
isopropyl alcohol. Refer to the "Care and Maintenance" section of this operator's manual.
Use a detergent that is safe for skin and washable surfaces. Most detergents can be high sudsing, so use
sparingly. Wipe with a damp, detergent free cloth to remove residue.
After cleaning the multiple use wristband, it should only be applied to the same patient; do not apply it to a
different patient.
Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling
of the device and device components, including batteries.
In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC,
do not dispose of this product as unsorted municipal waste. This device contains WEEE materials; please
contact your distributor regarding take-back or recycling of the device. If you are unsure how to reach your
distributor, please call Nonin for your distributor's contact information.
This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin.
Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the
following:
- excessive ambient light
- excessive motion
- electrosurgical interference
- blood flow restrictors (arterial
catheters, blood pressure cuffs,
infusion lines, etc.)
- moisture in the sensor
- improperly applied sensor
When using the monitor in the home, avoid exposing the monitor to lint and dust.
When using the monitor around small children and pets, avoid leaving the monitor unattended. Cables pose a
risk of injury, including strangulation.
Do not perform any testing or maintenance on this device while it is being used to monitor a patient.
This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of
this device is not possible. Except to replace batteries, do not attempt to open the case or repair the electronics.
Opening the case may damage the device and void the warranty.
2
, Model 3150, in liquid or clean it with agents containing ammonium chloride or
- incorrect sensor type
- poor pulse quality
- venous pulsations
- anemia or low hemoglobin
concentrations
- cardiogreen and other
intravascular dyes
- carboxyhemoglobin
Indications for Use
- methemoglobin
- dysfunctional hemoglobin
- artificial nails or fingernail
polish
- residue (e.g., dried blood, dirt,
grease, oil) in the light path