Standard Serviceability Test

The following procedure can be used to check the functions and facilities of the ADTS
405. For further information refer to Section 6 - Reference and Specification.
In the following:
All key presses are highlighted in bold and shown as identified on the key-pad.
Key presses inside brackets e.g., [MORE], are soft key presses (i.e., function key
selections {F1 to F4} indicated on the display screen).
Procedure
Connection and power-on checks
Connect power to the unit and, if necessary, connect the hand terminal.
Make sure that the blanking caps are fitted to the Ps and Pt front panel outputs.
Check power indicator is on.
Set the front panel switch to OPERATE. Set the front panel switch for the pump
unit to PUMP ON and check that the pump power indicator lights.
Check the display shows the first stage power-up message as below (where: x is
the version number of the installed display driver software):
Check the second stage power-up message displays as below (where: x is the
version number of the installed main operating software):
Note: "ADTS 405" may be replaced by an alternative identity string. The software version may
also be supplemented by extra alpha characters.
K114 Issue No. 9
DK 282 Version x.xx
Display Power up
Please Wait
DRUCK
ADTS 405
DK 263
VER 6.xx

Testing and Fault Finding