6.5 Specification

Physical Specification
Size
ADTS 405 ....................................... Standard 19" rack front panel 6U high (10½")
........................................................................ Depth behind front panel 255 mm (10")
.................................................................................... Rear panel connections available
ADTS 405F ............................... 442 mm x 755 mm x 332 mm (17½" x 30" x 13")
Weight
ADTS 405 ........................................................................................................ 12 kg [26.5 lb]
ADTS 405F .......................................................................................................... 29 kg [64 lb]
Power
......................................................................... Single phase AC in the range 90-260 V
............................................................................ Over the frequency range 47-440 Hz
........................................................................... No switching required - all automatic
DC power option ................................................................................................. 24 to 30 V
ADTS 405 ....................................................................................................................... 400 VA
ADTS 405F ..................................................................................................................... 500 VA
Pressure Fittings
Ps ............................................................................................................................. MS33656-6
Pt ............................................................................................................................. MS33656-4
Pressure supply ................................................................................................ MS33656-4
Vacuum supply ................................................................................................. MS33656-6
Pressure/Vacuum Requirements
The ADTS 405 requires a single vacuum inlet and a single air pressure inlet. At the control set-
point operating into a leak tight system, no air is consumed. A single vacuum source is
adequate for the rates detailed into the volumes listed with no crosstalk between Ps and Pt
channels. A pump capable of 14 litres/min (0.39 cubic ft/min) is required.
The air pressure source should be clean dry air at a pressure of 3 to 6 bar (45 to 90 psi) with
the ADTS 405 having an internal regulator to protect the controllers. Filtering should be to
better than 15 microns and a flow rate equivalent to 2.8 litres/min (0.1 cubic ft/min) at the
above pressure.
Note: The Druck PV 103 is a 19" rack unit that meets the requirements of the ADTS 405.
The integral pump rack provides all the requirements of the ADTS 405F.
Warm-up
If the ADTS 405 temperature has stabilized within the operating temperature range
before switching on, the ADTS 405 requires a warm-up period of 15 minutes to achieve
the stated accuracy.
6 - 35
K114 Issue No. 9