Druck ADTS 405 User Manual
6 - 39
Repeatability
Repeatability is defined as the change or variation in reading when the same pressure is
6
reapplied after an excursion between minimum and maximum of the range (i.e., including
hysteresis). The repeatability does not include curve fit errors.
Accuracy
Accuracy figures stated include curve fit errors, repeatability and temperature effects over the
range 5°C to 35°C (+41° to +95°F). The ADTS 405 can achieve the stated accuracy after a 15
.
minute warm-up period
K114 Issue No. 9