6 - 14
ENTER
The ENTER key completes numeric entry.
CLEAR/QUIT + ENTER (ABORT)
Pressing CLEAR/QUIT and ENTER together, restarts the ADTS 405 from the power-up
sequence.
Output valves close immediately and open during the restart sequence.
Note:
The restart sequence starts with the display showing:
The (ABORT) function should only be used as a last resort:
When a restart is required after the system has shutdown displaying an error message.
As an emergency measure with a serious problem concerning the aircraft or
component under test.
After this restart sequence, the system vents, at a safe rate, both channels (pitot and
static) to atmosphere during the power-up sequence.
K114 Issue No. 9
Reference and Specification
DRUCK
ADTS 405
DK127
Iss 5.xx