4.3

Maintenance Tasks

A
Visually inspect the unit for obvious signs of damage and check that all the equipment
is present; record any deficiencies.
B
Visually inspect the pneumatic output connectors for damage. Inspect the small o-ring
on each pneumatic output connector for cuts and any signs of wear; if necessary,
replace as detailed in 4.4. Visually inspect pneumatic hoses, electrical cables for cuts,
splits and damage; replace as necessary.
C
Before use, power-up the unit as detailed in Section 3. Check the date of the last
calibration and, if necessary, refer to task E. Record any error messages and refer to
Section 5.
D
Daily and before use, carry out the SST detailed in Section 5.
E
Normal calibration period 12 months. When calibration is due, the unit should be
withdrawn from service and returned to depot or calibration facility. The date of
calibration is stored in the unit and displayed in the power-up sequence and on a label
on the front panel. Do not use a unit with out-of-date calibration. Engineering
authority may change the periodicity of calibration depending on usage and the
operating environment.
F
As detailed in fault finding or when detailed after inspection replace the listed item in
table 2-1.
G
Clean the unit every week. Clean the front panel with a damp lint-free cloth (item 1,
Table 4-2) and mild liquid detergent (item 3, Table 4-2). Remove any stubborn dirt
using isopropyl alcohol (item 2, Table 4-2) and allow to dry. Clean all pneumatic
connectors with a soft brush (item 4, Table 4-2). Remove any stubborn dirt using
isopropyl alcohol (item 2, Table 4-2) and allow to dry.
H
Withdraw the unit from service and return to the repair depot. The intermediate
service manual contains the 1,000 hour servicing procedure.
J
Withdraw the unit from service and return to the repair depot. The intermediate
service manual contains the 3,000 hour servicing procedure.
4 - 3
K114 Issue No. 9