To apply an altitude of 5000 ft at a rate of climb of 6000 ft/min and an airspeed of 300
kts at a rate of 600 kts/min press the following keys:
SPEED/Qc then RATE to select rate of change of speed.
6, 0, 0, ENTER to set the rate.
SPEED/Qc to select airspeed.
3, 0, 0, ENTER to set an airspeed of 300 kts (airspeed [CAS] now starts increasing).
ROC/Ps Rate to select rate of climb.
6, 0, 0, 0, ENTER to set the rate of climb.
ALT/Ps to select the altitude.
5, 0, 0, 0, ENTER to set altitude (altitude [Alt] now starts increasing).
SPEED/Qc to view altitude and airspeed together.
Note:
When altitude and airspeed are changing at the same time, and automatic
airspeed rate is enabled, the system automatically adjusts the airspeed rate so
that the aim points are reached at the same time. The airspeed rate will not
exceed the entered aim value.
Wait for the aim values to be achieved.
Observe over a period of 1 minute that the value of Alt stays within ±10 ft and
the value of CAS with ±1 kt.
F
3-3 M
P
IGURE
AIN
RESSURE
D
(C
M
ISPLAY
ONTROL
ODE
3 - 13
)
K114 Issue No. 9
3