If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the
Pt remains unchanged*.
*
The display shows
SETUP, [UNITS] can be used to change the units of display, if set-up mode is not
switched off.
Before entering a new CAS or Qc aim, the rate of change should be set using the RATE
key.
In Control Mode (AIM display), a new CAS or Qc can be entered using the numeric keys.
Use LEAK MEASURE/CONTROL to change between Control and Leak Measure mode.
MACH/Pt
In aeronautical units (ft/kts or m/km/h), this key selects a Mach display.
In pressure units (mbar, inHg etc.), this selects a total pressure (Pt) display.
If the altitude aim is changed, after the entry of CAS, the Mach and TAS aim values
change while the CAS value remains unchanged*.
If the altitude aim is changed, after the entry of TAS or Mach, the CAS aim value
changes while the TAS and Mach values remain unchanged*.
If the Ps aim is changed, after the entry of Qc, then the Pt aim value changes while the
Qc remains unchanged*.
If the Ps aim is changed, after the entry of Pt, then the Qc aim value changes while the
Pt remains unchanged*.
=
*
The display shows
symbol next to the parameter that remains unchanged.
Before entering a new Mach or Pt aim, set the rate of change of speed using the
SPEED/Qc key followed by the RATE key.
In Control Mode (AIM display), a new Mach or Pt aim can be entered using the numeric
keys, Use LEAK MEASURE/CONTROL to change between Control and Leak Measure
mode.
=
symbol next to the parameter that remains unchanged.
6 - 3
K114 Issue No. 9