Cleaning And Disinfecting The Acquisition Module; Pre-Cleaning Inspection And Functional Testing; Cleaning The Acquisition Module - GE MAC VU360 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Component
Reusable electrodes

Cleaning and Disinfecting the Acquisition Module

Proper cleaning and disinfecting prolongs the life of acquisition devices. Failure to
use the proper cleaning solutions or to follow proper procedures can result in the
following:
• Appearance of waveforms when not connected to a patient, resulting in false
alarms instead of lead failure alarms
• Brittle and cracked device case
• Melting, dulling, or distortion of the case
• Total device failure, requiring replacement
• Unit malfunction
• Voided warranty
Soiled devices must be separated from non-contaminated devices to avoid
contamination of personnel or surroundings.

Pre-Cleaning Inspection and Functional Testing

Inspect the Acquisition Module each time you clean it or if you suspect a problem.
• Check the pins that the leadwires plug into. They should not be bent or loose.
Contact GE Healthcare Customer Support for any repairs needed.
• Check the Acquisition Module plastic case for any damage. Contact GE Healthcare
Customer Support for any repairs needed.

Cleaning the Acquisition Module

1.
Dispense Super Sani-Cloth
2.
If soil is present, thoroughly wipe the surfaces of the device with a fresh Super
Sani-Cloth
have been visibly removed.
Treated surfaces must remain visibly wet for a minimum of two minutes. Use
additional fresh disinfectant wipes, if needed, to ensure continuous two minutes
contact time. Pay attention to the recessed areas and ridges; use a cotton swab
to press onto the wipe for scrubbing these areas.
3.
Inspect the device and trolley to ensure the complete removal of soil from
surfaces, holes, and movable parts.
If soil is still present, re-clean the equipment by repeating step 2.
4.
Allow the device to air dry.
2088531-370-2
Visual Inspection
Refer to the supplier's instructions for reusable electrode cleaning,
disinfection and sterilization.
®
Wipe(s) from the canister.
®
Wipe for a minimum of two minutes and until soil and organic matter
  MAC VU360
Resting ECG Analysis System
Cleaning
Disinfection
Cleaning and Disinfection
 115

Advertisement

Table of Contents
loading

Table of Contents