Using Automatic
Calibration (cont.)
Advanced Programming
If you set the 90 minute Auto-Cal interval at 6:10 p.m., the next
Auto-Cal will occur at 7:30 p.m. (excluding the Auto-Cal performed
when first leaving the Autocal Menu).
If you enter a time interval that is not evenly divisible into 1440
minutes (1 day), the MMS-3 rounds up to the next acceptable interval.
You may now do one of the following:
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
[RUN]
Exit by pressing the
key.