Entering Saturation
Constants
4-12
To enter a saturation constant, you must enter 1 to 6 data points to
represent a curve of Cs (saturation constant) versus temperature. (For
more information, see Entering a Saturation Constant in Chapter 3.)
PanaView enables you to enter a saturation constant curve for each
channel. To enter a curve:
1. From the "New Meter Browser" option, click on the desired
communications port, meter, and channel.
2. From the "Edit" menu, select the "Properties" option. The
"Constant Saturation Table" window appears, as shown in
Figure 4-10 below.
3. Enter up to six data points to establish a curve of saturation
constants versus temperature.
4. The window also allows you to specify whether you wish
Enhanced Response for the channel. Enhanced Response uses a
dynamic moisture calibration technique to extrapolate the
moisture level to the end point when making measurements in
abrupt "dry down" conditions. (For more information, refer to
Using Computer Enhanced Response in Chapter 3.) Click the
appropriate radio button to turn Enhanced Response on or off.
5. When you have finished entering values, click on
the table immediately, and on
the window. To return the window to its previous settings, click on
[Cancel]
the
button.
Figure 4-10: Constant Saturation Table
[Apply]
[OK]
to confirm the table and close
Programming with PanaView