Entering A Saturation Constant - GE Moisture monitor series 3 Programming Manual

Panametrics hygrometer
Hide thumbs Also See for Moisture monitor series 3:
Table of Contents

Advertisement

March 2008
Entering a Saturation
Constant
3-28
The MMS-3 requires a saturation constant in order to calculate parts
per million by weight (ppm
saturation constant of the process is unknown, contact GE Sensing.
(Saturation constants are necessary for ppm
liquids only. See Appendix A of the Service Manual.)
To enter a saturation constant, the user must enter 1 to 6 data points to
represent a curve of Cs (saturation constant) versus temperature.
Enter curves for each channel separately.
[PROG]
Press the
key to enter the user program.
Enter Passcode: XXXX
Note: If you have already entered the user program, see Figure 3-2
on page 3-48 to navigate to the User Menu.
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
key to select the desired channel.
Programming Menu
[USER]
LOGGER
User Menu
[SAT-CONST]
#
T(°C)
Cs
1
+56
0
#
T(°C)
Cs
1
+56
33
#
T(°C)
Cs
1
+56
33
) in non-aqueous liquids. If the
w
in Henry's Law type
w
Enter the passcode
1
Use the arrow keys to move the
brackets to USER and press
1
Use the arrow keys to move the
brackets to SAT-CONST and
[YES]
press
.
1
If you want to begin entering
data at the data point displayed,
press the right arrow key to
move the bracket to T/°C.
Otherwise, enter the desired data
point number. Press [YES] to
confirm the entry. Then press the
right arrow key.
1
Enter the desired temperature.
[YES]
Press
to confirm the entry,
and press the right arrow key.
1
Enter the corresponding
saturation constant. Press
confirm the entry.
Advanced Programming
.
[CHAN]
[YES]
.
[YES]
to