Table 3-10: Measurement Modes, Required Inputs and Configuration Information
To measure:
Use these inputs:
Temperature
RH
and moisture
Moisture and
PPMv
pressure
Moisture and
PPBv
pressure
Moisture,
temperature
PPMw
and saturation
constant
Moisture and
MCF/IG
pressure
Moisture and
MCF/NG
pressure
Moisture and
PPMv/NG
pressure
Advanced Programming
Provided by:
TF Probe
M Series Probe w/ temp. option
M Series Probe & constant
temperature
TF Probe w/ pressure option
M Series Probe and auxiliary
pressure input
M Series Probe and pressure
constant
TF Probe w/ pressure option
M Series Probe and auxiliary
pressure input
M Series Probe and pressure
constant
TF Probe w/ temperature
option and saturation constant
M Series Probe w/ temperature
option & saturation constant
TF Probe, constant temp. and
saturation constant
M Series Probe, constant temp.
and saturation constant
TF Probe w/pressure option
M Series Probe and auxiliary
pressure input
M Series Probe and pressure
constant
TF Probe w/ pressure option
M Series Probe and auxiliary
pressure input
M Series Probe and pressure
constant
TF Probe w/ pressure option
M Series Probe and auxiliary
pressure input
M Series Probe and pressure
constant
March 2008
(see page 3-35)
Also configure the channel for:
TF (H, T, & P)
Mxx (H) and Kt (T)
Mxx (H & T)
TF (H, T & P)
Mxx (H) and X1 (P)
Mxx (H) and Kp (P)
TF (H, T & P)
Mxx (H) and X1 (P)
Mxx (H) and Kp (P)
TF (H & T) & refer to page 3-28
for saturation constant
Mxx (H & T) & refer to page 3-28
for saturation constant
TF (H), Kt (T); refer to page 3-28
for saturation constant
M (H), Kt (T) & refer to page 3-28
for saturation constant
TF (H, T & P)
Mxx (H) and X1 (P)
Mxx (H) and Kp (P)
TF (H, T & P)
Mxx (H) and X1 (P)
Mxx (H) and Kp (P)
TF (H, T & P)
Mxx (H) and X1 (P)
Mxx (H) and Kp (P)
3-37