Changing Alarm Settings
(cont.)
Programming with PanaView
5. To determine the alarm values, enter the Trip and Deadband
values in the designated text boxes on the window. (For a
definition of the Deadband value, see Setting Up the Alarms in
Chapter 3.)
6. To determine how the alarm will react when it encounters an error:
a. Pull down the "High" list box under the "On Error" line.
b. Then click on the desired action from the listing: "Ignore
error," "Trip on error," "Reset on error," or "Hold on last
error."
c. Repeat the procedure for the "Low" list box.
7. To test the alarm, click on the
button to trip the alarm, and on the
8. Click on
[Apply]
to immediately apply the changes, and on
confirm the changes and close the window. (To return the window
to its previous settings, click on the
[Test]
button. Click on the
[Reset]
button to reset it.
[Cancel]
button.)
March 2008
[Trip]
[OK]
to
4-5