Key
M5
Screw (M5)
Connector (Black, 3P)
Connector (White, 2P)
Electric lock
Plate meeting
Connector (Black, 3P)
Connector (Black, 2P)
Connector (Black, 4P)
Connect connector for electric lock with
connector (White, 2P) and fit to the angle.
Fit the key to lower side of the door.
Note Reuse screws to fit each parts.
(2P)
Electric lock should be secured by new
screws (M5).
Recommended to fit electric lock first to
the screw holes as arrow pointed in the
left picture when you fit electric lock to the
unit.
Fit plate meeting to the places as left
picture
connectors.
Tie
Harness for connector (Black, 3P)
should be put thorough the tie.
Peel seal off from insulation and fit the
insulation back to original position.
- 89 -
shown
and
connect
each