4. Various calibrations
<Temperature calibration Temp/CO2/O2 Calibration >
1. In Top Screen, press MENU key and select "Tools".
Press ENTER key to display "Select Tools (1/2)" display.
2. Select "Temp/CO2/O2 Calibration" and press ENTER key to display "Temp/CO2/O2 Calibration (1/2)".
Input actual measured value in the right side of "Temp Span" column.
3.
Press MENU key and select "OK". Press ENTER key to start temperature calibration automatically.
4.
Repeat the calibration until display temperature corresponds to the actual temperature.
Note Input range per calibration is within ±1.0
Ex. When set temperature is 37.0
0
Place temperature sensor in the center of the chamber and leave the apparatus until it is
stabilized.
1
In Top Screen, press MENU key and select "Tools" and press ENTER key.
2
Select "Temp/CO2/O2 Calibration" in "Select Tools (1/2)" display.
3
Press MENU key and select "OK" and press ENTER key.
4
Input "36.5" in the right side of "Temp Span" column in the "Temp/CO2/O2 Calibration (1/2)"
display.
5
Press MENU key and select "OK" and press ENTER key.
"Select Tools
6
against display value.
and actual temperature is 36.5
Procedures
1/2 " display is appeared.
- 32 -