Lighting The Uv Lamp For 24 Hours - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

The UV timer can be set from 0 to 30 min. The default setting is for 5 min. When the UV timer is set to
zero, the UV lamp will not light.
If the outer door is opened while the UV lamp is lit, the lamp will turn OFF. Then, when the door is
closed, the lamp will light for the preset period of time.
If the UV lamp ON time is set for longer than 10 minutes or if only the outer door is repeated opened
and closed, condensation may occur in the chamber and may affect temperature distribution. It will also
shorten the service life of the UV lamp.
When replacing the UV lamp, consult with a Sanyo representative or agent.
Note: To compensate for the drop in UV ray output along with increased accumulated ON time of UV
lamp, the Incubator automatically extends the UV lamp ON time according to the accumulated ON time.
The automatically extended time is displayed as a percentage under UV Timer Ext. in the UV Setting
Screen.

Lighting the UV lamp for 24 hours

If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
decontaminate the chamber by lighting the UV lamp for 24 hours.
1. Remove all attachments from the chamber, including the trays, tray supports, side supports, fan cover,
duct, fan, humidifying pan, humidifying pan cover, N
attachments in an autoclave or with disinfectant alcohol.
2. Clean and wipe off inside the chamber with disinfectant alcohol.
3. Set the CO
density to 0% and O
2
U V
S e t t i n g
1
U V
T i m e r
3
U V
L i f e
U V
T i m e r
4
5
U V
2 4 h
6
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key. The parameter will
be saved.
・This procedure must be preformed only with the outer door closed and the UV lamp turned OFF.
・The UV lamp will light continuously for 24 hours after this parameter is set. The setting is canceled if the
outer door is opened. Perform the above procedure to set the 24-hours lighting mode again if you open
the door.
・Install the attachments after completion of the 24-hour lighting mode.
Note: Set the upper limit alarm temperature to at least 10
when using the 24-hour lighting mode of UV lamp. The 24-hour lighting mode may cause an automatic
chamber temperature alarm because the temperature of the chamber will increase.
37
density to 20%. Set the UV 24h Mode to 1 on the UV Setting Display.
2
5
m i n
0
%
E x t
+ 0
%
M o d e
1
( 0 . O F F
- 144 -
or O
gas injection nozzle, and. Clean all the
2
2
( 0 - 3 0 m i n )
( 1 0 0 % = 1 0 0 0 h )
1 . O N )
o
C higher than the chamber set temperature

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)

Table of Contents