Control Specifications - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Control specifications

1.
Temperature control
(1) Setting range: 0.0 ~50.0 (Default setting: 37.0 )
(2) Automatic set temperature alarm
High temperature alarm setting range: Set temperature +1.0~+5.0 (Default setting:+1.0 )
Low temperature alarm setting range: Set temperature -1.0~-5.0 (Default setting:-1.0 )
(3) Alarm delay time: 0~15 minutes (Default setting: 15 minutes)
(4) How to set temperature
In TopScreen, press MENU key and select Set to display Stand-by Setting screen.
Move a curser on "Temperature" and set a value by pressing numeric keys.
Press MENU key and select OK and press ENTER key to memorize the value.
2.
CO2 control
(1) Setting range: 0.0%~20.0%(Default setting: 0.0%)
(2) Automatic set CO2 density alarm
High CO2 density alarm setting range: Set density +0.5%~+5.0% (Default setting: +1.0%)
Low CO2 density alarm setting range: Set density -0.5%~-5.0% (Default setting: -1.0%)
(3) Alarm delay time: 0~15 minutes (Default setting: 15 minutes)
(4) How to set CO2 density
In TopScreen, press MENU key and select Set to display Stand-by Setting screen.
Move a curser on "CO2 Density" and set a value by pressing numeric keys.
Press MENU key and select OK and press ENTER key to memorize the value.
3.
N2/O2 control
(1) Setting range: 1.0%~80.0% (Default setting: 20.0%)
It is uncontrolled O2 density when O2 setting range is exceeding 18.0% and less than 22.0%.
(2) Automatic set O2 density alarm
High O2 density alarm setting range: Set density +0.5%~+5.0% (Default setting: +1.0%)
Low O2 density alarm setting range: Set density -0.5%~-5.0% (Default setting: -1.0%)
(3) Alarm delay time: 0~30 minutes (Default setting: 30 minutes)
(4) How to set O2 density
In TopScreen, press MENU key and select Set to display Stand-by Setting screen.
Move a curser on "O2 Density" and set a value by pressing numeric keys.
Press MENU key and select OK and press ENTER key to memorize the value.
4.
Alarms, safety functions and self diagnosis functions
(1) Upper limit temperature alarm
Setting range: Approx.35 ~51
When a chamber temperature is over the setting upper limit temperature, OVERHEAT indicator
will illuminate, buzzer will sound continuously and remote alarm will be active.
- 16 -

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)

Table of Contents