<Power failure during H2O2 decontamination mode>
If power fails while H2O2 misting process (Step 4) or UV resolution process (Step 7), UV resolution
will reset and repeat.
If power fails while H2O2 misting process (Step 4), "Decon Failed" will be displayed and "Err35: Power Failed"
will be displayed on the message bar.
If power fails while UV resolution process (Step 7), "Decon Failed" will be displayed and "Err37: Power Failed"
will be displayed on the message bar.