Sanyo MCO-19M Service Manual page 44

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Ex UV lamp illumination time is set in 5 minutes;
0
In Top Screen, press MENU key and select "Tools".
1
Press ENTER key and select "UV Setting" column in "Select Tools (1/2)" display.
2
Press MENU key and select "OK".
3
Press ENTER key and input "5" in the right side of "UV Timer" in "UV Setting" display.
4
Press MENU key and select "OK".
5
Press ENTER key and "Select Tools (1/2)" display is appeared.
<UV lamp accumulation time>
Display range of UV lamp accumulation time is 0~260%.
Note) UV lamp accumulation time is displayed by percentage. 1000Hours = 100%
<UV lamp 24hours illumination mode>
Chamber wall will be sterilized by UV lamp in this mode.
When you input "0" in the right side of "UV 24h Mode" column in "UV Setting" display, UV lamp 24hours illumination
mode is inactive.
When you input "1" in the right side of "UV 24h Mode" column in "UV Setting" display, UV lamp 24hours illumination
mode is active.
Note) If you open an door during UV lamp 24hours illumination mode, UV lamp 24hours illumination mode will be
forcibly inactive on the way.
The detection for UV lamp malfunction will not be done during UV lamp 24hours illumination mode.
Ex) When you set UV lamp 24hours illumination mode;
0
In Top Screen, press MENU key and select "Tools".
1
Press ENTER key and select "UV Setting" column in "Select Tools (1/2)" display.
2
Press MENU key and select "OK".
3
Press ENTER key and input "1" in the right side of "UV 24h Mode" in "UV Setting"
display.
4
Press MENU key and select "OK".
5
Press ENTER key and "Select Tools (1/2)" display is appeared.
(UV lamp relay will be active and UV lamp will illuminate)
Procedures
Procedures
- 41 -
MENU key
Tools
OK
OK
MENU key
Tools
OK
OK