How to install UV lamp add-on kit (MCO-19UVS)
Kit includes:
1 UV lamp
1 UV lamp cover
6 Screws M4
1. Unplug the unit and make sure that power is
not supplied to the unit.
2. Remote the duct from inside the unit, then
take 2 caps and 2 screws off as shown in
Fig.1
Cap
Fig. 1
3. Take the rear cover off as shown in Fig. 2.
Rear cover
2 Glow ass'y
1 Ballast
Screw
Fig. 2
4. Take 2 insulations out for the UV lamp lead
wire hole as shown in Fig. 3. Insulations can
be easily taken out by pushing it from inside
of the chamber. Keep them for reuse.
Insulation
5. Set the UV lamp ( ) to the hole and make sure
to fit correctly. Fix the UV lamp cover ( ) by 2
screws ( ).