Connection Of Co 2 Gas Cylinder - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Connection of CO
WARNING
When connecting a gas cylinder to the Incubator, confirm the gas type. Confirm that the connections
are secure and that no gas will escape. Be sure to use the specified pressure. Using an incorrect
gas or pressure may result in explosion or fire, or in gas poisoning or oxygen deprivation due to escaping
gas.
Install the Incubator in a location with adequate ventilation. If adequate ventilation cannot be
provided, then install an alarm system using
1. Use a liquefied CO
gas cylinder (at least 99.5% pure). The siphon (dip tube) type cannot be used.
2
2. Install the Gas Pressure Regulator (MCO-100L, purchased separately) specified by Sanyo on the CO
gas cylinder. Otherwise, use a CO
for the primary pressure, and 0.2 MPaG (2 kgf/cm
3. Using the gas supply pipe that is provided, connect the CO
inlet (located at the lower left-hand side on the rear panel of the Incubator).
Note:
If CO
is supplied to multiple CO
2
gas pressure regulator. The gas pressure regulator safety valve will operate, and there may an explosive
sound.
When the MCO-21GC is not mounted
Using the gas supply pipe that is provided, connect the CO
cylinder to CO
gas pipe connection port A on the Incubator. After connecting the pipe, check to make
2
sure that no gas is leaking.
When the MCO-21GC is mounted
Connect a pair of CO
gas cylinders with CO
2
will be switched automatically. Connect the main cylinder to port A and the reserve cylinder to port B. After
connecting the cylinders, check to make sure that no gas is leaking.
For details on installing the optional automatic CO
the MCO-21GC installation guide. For details on using the MCO-21GC, refer to page 67.
4. Set the CO
gas on the secondary side to 0.03 MPaG (0.3 kgf/cm
2
pressure increases, the CO
cause gas supply lines inside the Incubator to come loose, which may result in gas poisoning or oxygen
deprivation due to the escaping of gas. If gas lines come loose, the Incubator must be repaired.
The gas lines connected to the Incubator will degrade over time. If any deterioration or abnormalities
are found during inspection, replace the lines immediately.
gas cylinder
2
CO
and O
2
2
gas pressure regulator rated at 25 MPaG (250 kgf/cm
2
2
G, 30 psiG) for the secondary pressure.
Incubators from a single gas cylinder, a CO
2
gas pressure regulators independently. The gas supply line
2
Cylinder Changeover System (MCO-21GC), refer to
2
gas concentration control range will increase. Excessive pressure may
2
- 123 -
densitometers.
gas pressure regulator to the CO
2
solid will be formed in the
2
gas pressure regulator for the CO
2
2
G, 4.3 psiG) for gas injection. As the
2
2
G, 3600 psiG)
gas pipe
2
gas
2
16