Sanyo MCO-19M Service Manual page 149

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

3. The Date Time Display will appear. Set the date, time and log interval.
D a t e
1
3
D a t e
T i m e
4
5
L o g
I n t e r v a l
6
・Entering the date
Example for October 1, 2010: Enter 101001 in the Date Field.
・Entering the time
Example for 10:05:00: Enter 100500 in the Time Field.
・Entering the log interval
Example for 6 minutes: Enter 6 in Log Interval Field.
Note:
The default is 6 minutes.
The log interval can be set to between 2 and 30 minutes.
Relation between the log interval and the period that can be saved
1: Log interval of 2 minutes -- About 5 days
2: Log interval of 6 minutes -- About 14 days
3: Log interval of 30 minutes -- About 70 days
For the data beyond the above period, the older data is deleted to save the newer data.
39
T i m e
1 0 / 1 0 / 0 1
( Y Y / M M / D D )
1 0 : 0 5 : 0 0
( h h : m m : s s )
6 m i n
( 2 - 3 0 m i n )
- 146 -