Sanyo MCO-19M Service Manual page 83

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Removal of MCO-HP on the way of decontamination
#
1
Turn the unit off and on.
2
When 'Warming up' is displayed, press
MENU key and select 'about' and press
ENTER key.
3
Remove MCO-HP and press MENU key
and ENTER key.
4
Ensure electric lock should be unlocked.
(3) Automatic CO
Density Calibration Add-on Kit (MCO-SG)
2
Setting of Standard CO
#
1
2
Press MENU key.
3
Select Tool and press ENTER key.
4
Select "Standard Gas setting" on page 2.
Press ENTER key.
5
Input standard gas density by numeric
keys.
6
Press MENU key.
7
Select 'OK' and press ENTER key.
CO
and O
density calibration
2
2
#
1
2
Press MENU key.
3
Select Tool and press ENTER key.
4
Select "Standard Gas calibration" on the
nd
2
page of "Select Tool" screen.
Press ENTER key.
5
Select 'OK' in "MENU" window and press
ENTER key to start CO
6
Operation
and O
density
2
2
Operation
Operation
calibration.
2
Key
Power switch
MENU
To avoid the device turning back
ENTER
to resolution process, ensure to
press "MENU" key and select
'about' and press "ENTER" key
when 'Warming up' is displayed.
MENU
ENTER
Ensure outer door should be
opened.
Key
-----
MENU
ENTER
ENTER
"Standard Gas Setting" screen
Numeric key
MENU
ENTER
Key
-----
MENU
ENTER
ENTER
"Standard Gas Calibration" screen
(System check starts)
After
confirmation screen is shown to
start calibration.
ENTER
After CO
"Top Screen" is shown.
- 80 -
Display/Remark
"H2O2 Decontamination"
"Top Screen"
Display
"Top Screen"
"MENU" window
"Select Tool"
"MENU" window
Display/Remark
"Top Screen"
"MENU" window
"Select Tool"
passing
system
check,
calibration is complete,
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)

Table of Contents