Setting Procedures For Optional Components - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Setting procedures for optional components

1. How to access to optional components
(1) Input of service code
1
2
Press MENU key.
3
Select "Tool" and press ENTER key.
4
Press "MENU" key.
5
Select "SVC" and press ENTER key.
6
Input service code '384' by numeric keys.
7
Press MENU key.
8
Select OK and press ENTER key.
Note) If you fails to input service code, "Wrong Password" is displayed under the "Password" on the
screen and error tone emits. The password "384" should be input correctly.
Service code is unchangeable.
(2) Access to optional components
#
1
2
Select "System Configuration" and press
ENTER key.
3
Select applied optional components* from
"System Configuration" and input "1" by
numeric key.
4
Press MENU key.
5
Select "OK" and press ENTER key.
* Optional components in "System Configuration"
Optional components
UV lamp add-on kit (MCO-19UVS)
H
O
generator (MCO-HP),
2
2
H
O
Decontamination kit (MCO-HL)
2
2
Automatic CO
Density Calibration Add-on Kit (MCO-SG)
2
Automatic CO
Cylinder Changeover System (MCO-21GC)
2
Applicable for; MCO-19UVS, MCO-HL, MCO-HP, MCO-SG
Operation
Operation
Key
-----
MENU
ENTER
MENU
Numeric key
MENU
ENTER
Service mode "Select Tools Svc"
Key
-----
Service mode "Select Tools Svc"
ENTER
"System configuration" window
Numeric key
MENU
ENTER
Application in "System Configuration"
"CO2 Gas AutoChanger"
- 77 -
Display
"Top Screen"
"MENU" window
Select "Tool" screen
"MENU" window
Password input screen
"MENU" window
Display
"MENU" window
"MENU" window
"UV Function"
"Decon Function"
"Std Gas Cal Function"

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)

Table of Contents