Specifications Of Sensors - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

Specifications of sensors

Temp. sensor, Overheat protect sensor, CO
Temperature
Resistance
( )
(k )
0
27.28
1
26.13
2
25.03
3
23.99
4
22.99
5
22.05
6
21.15
7
20.29
8
19.48
9
18.70
10
17.96
11
17.24
12
16.55
13
15.90
14
15.28
15
14.68
16
14.12
CO
sensor (Type: IR sensor)
2
CO
density
Output voltage
2
V
0.0
4.00
1.0
3.82
2.0
3.63
3.0
3.49
4.0
3.39
5.0
3.31
6.0
3.25
7.0
3.19
BOX temp. sensor (Type:103AT-1)
2
Temperature
Resistance
( )
(k )
17
13.57
18
13.06
19
12.56
20
12.09
21
11.63
22
11.20
23
10.78
24
10.38
25
10.00
26
9.63
27
9.28
28
8.94
29
8.62
30
8.31
31
8.02
32
7.73
33
7.46
CO
density
Output voltage
2
V
8.0
3.14
9.0
3.10
10.0
3.06
11.0
3.02
12.0
2.99
13.0
2.96
14.0
2.93
15.0
2.90
*Reference value for R40 in Main PCB
- 72 -
Temperature
Resistance
( )
(k )
34
7.19
35
6.94
36
6.70
37
6.47
38
6.25
39
6.03
40
5.83
41
5.63
42
5.44
43
5.26
44
5.08
45
4.91
46
4.75
47
4.59
48
4.44
49
4.30
50
4.16
CO
density
Output voltage
2
V
16.0
2.88
17.0
2.85
18.0
2.83
19.0
2.81
20.0
2.78

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19m(uv)

Table of Contents