How To Install Automatic Co 2 Density Calibration Add-On Kit Mco-Sg - Sanyo MCO-19M Service Manual

Multi-gas incubator
Hide thumbs Also See for MCO-19M:
Table of Contents

Advertisement

How to install Automatic CO2 density
calibration add-on kit MCO-SG
Parts descriptions
Plate meeting
Sensor BOX
Rear cover
Sensor BOX
(5) Plate meeting
(3) Label
(6)Clip
(4) Filter
(1) Tube joint
Tube joint
Clip
Label
Filter
Tube connect
Cut 30mm off from the tube
Tube connect
(2) Clip
(Put into the hole)
Connector (White 5P)
Connector (White, 2P)
Clip
Screw (M4×10)
Remove rear cover and sensor BOX
cover.
Cut 30mm off from the tube and insert
tube connect to the place as the left
picture specified.
Follow the procedure (1)~(6) to fit each
parts.
(1) Fit the tube joint.
(2) Apply clips to the holes.
(3) Paste the label between clips.
(4) Fit the filter. Make sure the orientation.
(5) Fit the plate meeing.
(6) Apply clips as left picture specified.
Place the tube as the left picture specified.
Note Take care for the tube should not be
bent.
Connect each parts with connectors.
- 101 -
Tools
Philips screwdriver
Monkey wrench