modes may be used onboard the aircraft without any further testing by the operator. This device can
withstand exposure to common electrostatic (ESD) and electromagnetic interference (EMI).
Guidance and Manufacturer's Declaration –
Electromagnetic Immunity
The transmitter (P/N 9438-06) is intended for use in the electromagnetic environment specified
in the next table. The customer or the user of the transmitter should ensure that it is used in such
an environment.
Transmitter Electromagnetic Immunity Specifications
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
Receiver Product Specifications
Part Number
Reading Frequency
Dimensions
TX/RX Frequencies
Bandwidth
Maximum Output Power
Modulation
Data Rate
Dexcom G5 Mobile System User Guide
Technical Information
IEC 60601 Test
Level
± 8 kV Contact
± 15 kV Air
30 A/m
(Continued on next page)
Transmitter
Electromagnetic
Compliance
Environment Guidance
Level
Floors should be wood, concrete or
± 8 kV Contact
ceramic tile. If floors are covered
with synthetic material, the relative
± 15 kV Air
humidity should be at least 30%.
Power frequency magnetic
fields should be at levels
30 A/m
characteristic of a typical location
in a typical commercial or hospital
environment.
MT22719
Every 5 minutes
Length: 4.0 inches
Width: 1.8 inches
Thickness: 0.5 inches
2.402-2.480 GHz
1.22 MHz
2.5 mW EIRP
Gaussian Frequency-Shift Keying
1 Mbps
277