or 20%, 30 mg/dL or 30% or greater than 40 mg/dL or 40% are provided in Table 1-A and 1-B. The
tables are categorized within CGM glucose ranges. When you see a CGM reading on your receiver,
this table shows you how likely that reading matches your blood glucose level (measured by YSI in
the study).
SW10505 Adult
For example, in the
Study (Table 1-A), the total number of data pairs considered in
the analysis was 2263. Of these, 93% of the System readings fall within ± 20 mg/dL of the YSI blood
glucose values ≤ 80 mg/dL and within ± 20% of YSI blood glucose values > 80 mg/dL.