Key Reference

VBW

Changes the analyzer post- detector filter from
3 Hz to 1 MHz in a 1, 3, 10 sequence using the
knob, the step keys, or the numeric keypad.
When an unavailable bandwidth entered using the numeric
N O TE
keypad, the closest available bandwidth in the 1, 3, 10
sequence is used.
As the video bandwidth is decreased, the sweep
time is increased to maintain amplitude calibration.
A "#"mark appears next to RBW on the display
when it is not coupled.
Key Access: [BW/SWP]

VBW/RBW

Selects the ratio between the video and resolution
bandwidths.
If signal responses near the noise level are visually
masked by the noise, the ratio can be set to less
than 1 to lower the noise. The knob and step keys
change the ratio in a 1, 3, 10 sequence.
When preset is set to default, or VBW/RBW is set to Auto,
N O TE
the ratio is set to 1.
Key Access: [BW/SWP]

Avg Type

Accesses the functions to automatically or
manually choose from one of the following
averaging type:
72
N9340A User's Guide